Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: Yokogawa HR2300: CAL ADJ /5520 DATE: 15-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 122 CONFIGURATION: Fluke 5520A ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # yo23_52.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- P J F W 1.002 ASK+ K X 1.003 HEAD PRELIMINARY INSTRUCTIONS 1.004 DISP Connect UUT to AC power and turn unit ON. 1.004 DISP 1.004 DISP Warm-up time: 1-hour - REQUIRED 1.004 DISP Temperature: 23 +/-2 degC 1.004 DISP Humidity: 55 +/-10% 1.005 DISP [32] WARNING 1.005 DISP HIGH VOLTAGE is used or exposed during the performance 1.005 DISP of this calibration. DEATH ON CONTACT may result if 1.005 DISP personnel fail to observe safety precautions. 1.006 HEAD CONNECTIONS 1.007 PIC yo23_52 1.008 OPBR Are Channels 1,2,3,4,5 & 6 connected properly? 1.009 JMPF 1.007 1.010 DISP Save settings to IC card or print "SET UP LIST" 1.011 DISP Turn UUT power off OFF. 1.011 DISP 1.011 DISP Set switch 1,2,4,8 of the DIP switch behind the chart 1.011 DISP cassette to ON. 1.012 DISP Hold LOWER DISPLAY key depressed and turn UUT on. 1.012 DISP 1.012 DISP Hold the LOWER DISPLAY key until 'SET UP MODE=RECORD' 1.012 DISP appears on the display. 1.013 DISP Select 'A/D' using function key and press ENTRY key. 1.013 DISP 'A/D MODE=DC DISP' appears on the display. 1.013 DISP 1.013 DISP Select 'DC AVE' using function key and press ENTRY. 1.013 DISP '20mV ZERO' appears on the display. 1.014 DISP Press ENTRY key. The UUT adjusts 0mV automatically 1.014 DISP and then 'A/D MODE=DC AVERAGE' appears on the display. 1.015 DISP Press ENTRY key. 1.015 DISP 'AVE=20mV ZERO' appears on the display 1.016 DISP Select '0ohm:1-Z' using function key and press ENTRY. 1.016 DISP 1.016 DISP The UUT adjusts 0 Ohms automatically and then 1.016 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.017 DISP Press ENTRY key. 1.017 DISP 'AVE=0ohm:1mA' appears on the display 1.018 ASK- U # Insure lead resistance is less than 10 Ohms. 1.019 5520 100.0Z S 2W 1.020 ASK+ U 1.021 DISP Select '100ohm:1' using function key and press ENTRY. 1.021 DISP 1.021 DISP The UUT adjusts 100 Ohm automatically and then 1.021 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.022 DISP Press ENTRY key. 1.022 DISP 'AVE=100ohm:1mA' appears on the display. 1.023 DISP Select '100:1H' using function key and press ENTRY. 1.023 DISP 1.023 DISP The UUT adjusts 100 Ohm automatically and then 1.023 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.024 DISP Press ENTRY key. 1.024 DISP 'AVE=100ohm:1mA-H' appears on the display 1.025 DISP Select '0ohm:2-Z' using function key and press ENTRY. 1.025 DISP 1.025 DISP The UUT adjusts 0 Ohm automatically and then 1.025 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.026 DISP Press ENTRY key. 1.026 DISP 'AVE=0ohm:2mA' appears on the display. 1.027 ASK- U # Insure lead resistance is less than 10 Ohms. 1.028 5520 100.0Z S 2W 1.029 ASK+ U 1.030 DISP Select '100ohm:2' using function key and press ENTRY. 1.030 DISP 1.030 DISP The UUT adjusts 100 Ohm automatically and then 1.030 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.031 DISP Press ENTRY key. 1.031 DISP 'AVE=100ohm:2mA' appears on the display. 1.032 DISP Select '100:2H' using function key and press ENTRY. 1.032 DISP 1.032 DISP The UUT adjusts 100 Ohm automatically and then 1.032 DISP 'A/D MODE=DC AVERAGE' appears on the display. 1.033 DISP Select 'WRITE' using function key and press ENTRY. 1.033 DISP 1.033 DISP 'WRITE OK=NO' appears on the display. 1.033 DISP 1.033 DISP Select 'YES' using function key and press ENTRY key. 1.033 DISP 'A/D MODE=WRITE' appears on the display. 1.034 DISP Select 'ESC' using function key and press ENTRY. 1.034 DISP '***A/D SET***' appears on the display. 1.035 DISP Press ENTRY key. 1.035 DISP 'SET UP MODE=A/D-ADJ' appears on the display. 1.035 DISP 1.035 DISP Select 'END' using function key and press ENTRY. 1.035 DISP 1.035 DISP The UUT returns to normal operation. 1.037 END